Customization: | Available |
---|---|
After-sales Service: | 2years |
Warranty: | 2years |
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
|
---|---|
Support payments in USD |
Secure payments: | Every payment you make on Made-in-China.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Quantity(units) | 1 - 10 | >10 |
Lead time (days) | 15 | To be negotiated |
Characteristics:
1.Direct displaying conductivity, resistivity, TDS and salinity
2.Support calibration (constant or TDS factor, up to two point calibration)
3.Power-off protection, protect your data
4.Remind the electrode performance and calibration
5.Auto upgrade hardware
6.Support GLP norm: More complete checking the stored information
Support view, print, delete stored data
It can record, view and print calibration data.
It can store measuring data(600 sets) which follows GLP norm.
The meter requires setting operator No and records the all operating procedures.
Technical data:
Model |
DDSJ-308F |
|
Parameters |
Conductivity/Resistivity/TDS/Salinity/ºC |
|
Range
|
Conductivity |
0.000μS/cm~199.9mS/cm (Full range measuring, no need to exchange electrode) |
Resistivity |
5.00Ω·cm~20.00MΩ·cm |
|
TDS |
0.000 mg/L~99.9g/L |
|
Salinity |
(0.00~8.00)% |
|
Temp |
-5.0~110.0ºC |
|
Accuracy |
Conductivity |
±0.5%FS |
TDS |
±0.5%FS |
|
Salinity |
±0.1% |
|
Temp |
±0.2ºC |
|
Stability |
±0.3%(FS)/3h |
|
Temp compensation |
Manual/Auto(-5.0~110.0)ºC |
|
Temp coefficient |
Adjustable |
|
Calibration |
May calibrate electrode constant or TDS factor |
|
GLP Standard |
Yes |
|
Measuring Mode |
3 (continuous measuring, timed measuring and smart measuring mode) |
|
Display |
LCD screen, backlight |
|
Power |
Universal power adapter (9V DC, 800mA) |
|
Communication Interface |
USB |
|
Dimension&Weight |
280×215×92mm, 1kg |
|
Standard Configuration |
DJS-1D Platinum Black Conductivity Electrode T-818-B-6 ATC Probe |